7.0
Repairs and Replacements
7.1
Safety Warning - Disconnection from Power Supply
Immediately switch the product off in the event of unforeseen
circumstances (e.g. large amount of smoke). Allow the product to return
to room temperature before inspection.
Always ensure that the product is disconnected from the electrical supply
before repair work is carried out.
Caution
: Double pole/neutral fusing may be used in this product.
7.2
Safety Warning - Refractory Fibre Insulation
Insulation made from High Temperature Insulation Wool
Refractory Ceramic Fibre, better known as (Alumina silicate
wool - ASW).
This product contains
alumino silicate wool
products in its thermal insulation. These
materials may be in the form of blanket or felt, formed board or shapes, slab or loose
fill wool.
Typical use does not result in any significant level of airborne dust from these materials,
but much higher levels may be encountered during maintenance or repair.
Whilst there is no evidence of any long term health hazards, it is strongly recommended
that safety precautions are taken whenever the materials are handled.
Exposure to fibre dust may cause respiratory disease.
When handling the material, always use approved respiratory protection
equipment (RPE-eg. FFP3), eye protection, gloves and long sleeved clothing.
Avoid breaking up waste material. Dispose of waste in sealed containers.
After handling, rinse exposed skin with water before washing gently with
soap (not detergent). Wash work clothing separately.
Before commencing any major repairs it is recommended to make reference to the
European Association representing the High Temperature Insulation Wool industry
(www.ecfia.eu).
Further information can be provided on request. Alternatively, Carbolite Gero Service
can quote for any repairs to be carried out either on site or at the Carbolite Gero
factory.
7.3
Panel Removal
Except for temperature controller replacement, access to components is by removal of
the top panel. Remove 10 self-tapping screws. Carefully save the 10 spacers (between
the product case and the top panel). Lift off the top panel; ensure there is somewhere
to place it without pulling on the wire connections.
31
7.0 Repairs and Replacements
Summary of Contents for CARBOLITE GERO VMF-ASTM
Page 4: ...10 2 Power Settings 38 11 0 Specifications 39 11 1 Environment 39 4 ...
Page 20: ...3 0 2132 Controller 3 7 Navigation Diagram 20 ...
Page 28: ...6 0 Maintenance Power Consumption Measure the current drawn on each phase cir cuit 28 ...
Page 40: ......
Page 41: ...Notes Service Record Engineer Name Date Record of Work ...