4.0
Operator Levels
There are 4 operator levels available within the 3016. These user levels control access
to functionality within the controller via permissions.
The default operator level is
Level 1
.
Level 1
Level 2
Level 3
Config. Level
Functionality
Permissions
Login / Change User Level
Access Home Display
Change Setpoint (including Over-Temperature Setpoint)
Run a Pre-configured Program
Edit / Create a Program
Edit Controller Settings
4.1
Changing Operator Level
4.1.1
Operator Level 2
To access Operator Level 2:
1. From the Home display, press and hold the page button for 3 seconds until the
GOTO display appears.
2. Use the up and down arrow buttons to select the operator level you wish to access.
3. Press the scroll button to confirm. The CODE display will appear. You now have to
input a passcode.
4. Use the up and down arrows to change the number input.
5. Press the scroll button to move to the next digit.
6. When you have finished inputting the passcode, wait for 2 seconds. If entered cor-
rectly, PASS will appear on the display and you will now have access to additional
functionality.
Note:
The passcode to access Operator Level 2 is
0009
8