previous segment in the program. Dependent on "Dwell
Units" parameter.
Maximum values if the "Dwell Units" are:
l
SECS
= 99:59
l
MINS
= 99:59
l
HRS
= 500.0
EV.OP
Note:
The 3016 is available with 2 optional relay outputs
which may be configured as event outputs or alarms.
Any parameters above
3
are not applicable to this
controller and should not be used. Do not use values
above the number of events configured. If the 3016 has
been ordered without the relay option, the EV.OP
parameter will not appear.
l
0
- All event outputs are OFF
l
1
- Only event output 1 is ON
l
2
- Only event output 2 in ON
l
3
- Both event output 1 and 2 are ON
StEP
TSP
Used to define the Target Setpoint for that segment i.e.
the temperature (°C) that is to be reached at that point in
the program.
EV.OP
Note:
The 3016 is available with 2 optional relay outputs
which may be configured as event outputs or alarms.
Any parameters above
3
are not applicable to this
controller and should not be used. Do not use values
above the number of events configured. If the 3016 has
been ordered without the relay option, the EV.OP
parameter will not appear.
l
0
- All event outputs are OFF
l
1
- Only event output 1 is ON
l
2
- Only event output 2 in ON
l
3
- Both event output 1 and 2 are ON
ENd
-
Used to trigger the end of the program. The behaviour of
the controller during at the end of a program is determined
by the "End Type" parameter.
6.4
Event Outputs - Customer Notes
The 3016 has eight available event outputs; connections must be specified at the time of
order.