Note:
Also adjustable at Operator Level 3.
FILT.T
Filter Time
Constant
A filter is provided to reduce the frequency of electrical
noise picked up by the controller, which may interfere with
its function. The higher the Filter Time Constant value, the
more the effects of electrical noise are reduced.
The default value is
1.6
.
Note:
The higher the Filter Time Constant value, the
slower the the measured temperature will be to respond
to any fluctuations. Only adjust this value if necessary.
Note:
Also adjustable at Operator Level 3.
SB.OUT
Sensor Break
Output
Indicates whether a sensor break has been detected. This
usually occurs when a thermocouple fails or is
disconnected.
l
OFF
- No sensor break detected
l
ON
- Sensorbreak detected
PV
Process Variable Displays the current actual temperature of the product.
PV .ST
Process Variable
Status
Displays the status of the Process Variable (actual
temperature).
If the PV.Status displays anything but
Good
, please contact
Carbolite Gero Service.
LOOP
List
s.list
Sublist
Use the up and down arrows to view sublists of additional
parameters. Use the scroll button to select a sublist.
l
MaIN
- Access parameters relating to Main Control
Loop
l
AtuN
- Access parameters relating to Autotune
mAIN
Sublist:
R-L
Remote-Local
Select
Used to set the source of the setpoint.
l
Loc
- The controller uses the setpoint input within the
controller
l
rEm
- The controller uses the setpoint from a remote
source e.g. another controller. This option is most com-
monly used in products with multiple heated zones, or
cascade control.
Note:
Also adjustable at Operator Level 3.
PV
Process Variable Displays the current actual temperature.
31