CENTRA – MDS/LDS/RDS US Operator Manual
ELGA
Page 40
CENTRA – MDS/LDS/RDS US Version 01 – 08/14
10.
SANITIZATION PROCEDURES
The unit is sanitized to help reduce the bacteria within the
pipework, and reservoir. Please read this entire section to become
familiar with the procedure before you start. It is recommended that
feed water supplies to the
CENTRA
are sanitized once a month to
maintain low bacterial counts. Additionally, recirculation sanitization
is recommended every 12 months. It should also be carried out:
•
If the unit has not been used for a prolonged period of
time.
•
If the unit is to be operated under adverse conditions e.g.
high temperature.
•
After replacement of filters or major maintenance.
WARNING! ALWAYS WEAR RUBBER GLOVES, APRON AND
FACEMASK. DO NOT BREATHE FUMES OR
ALLOW TO COME IN CONTACT WITH SKIN OR
EYES. ALWAYS FOLLOW RECOMMENDATIONS
FOUND IN THE MANUFACTURERS MATERIAL
SAFETY DATA SHEET AND ANY APPLICABLE
OSHA STANDARDS FOR THE CHEMICAL BEING
USED.
WARNING! LABEL THE MACHINE WITH APPROPRIATE
WARNING SIGNS SUCH AS “DO NOT USE/
CONTAINS STERILANT” (NOT PROVIDED)
Minncare Cold Sterilant is a Peracetic and Hydrogen Peroxide
based solution.
Refer to the Minncare label for additional information and follow all
applicable directions for use on the manufacturer’s label in
conjunction with the following instructions.
Selection of:
10.1
Automatic recirculation sanitization (RDS)
1.
ENSURE the process is OFF.
2.
PRESENT sanitization PASSkey (green key).
3.
PRESS TICK
.
10.2
Automatic recirculation sanitization procedure
(RDS)
1.
AUTOMATIC level adjustment in the reservoir will
occur (start point between 60 - 90 liters)
2.
The display will tell you if reservoir needs to be filled
or drained, PRESS process.
3.
When directed ENSURE that, if fitted, ion-exchange/
purification media cylinders are isolated by closing
isolating valve (V6) and opening bypass valve (V7).