CENTRA – MDS/LDS/RDS US Operator Manual
ELGA
Page 38
CENTRA – MDS/LDS/RDS US Version 01 – 08/14
13. PRESS PROCESS button and allow unit to enter
recirculation.
14. CHECK filter housing for leaks.
15. REPLACE spanner in door and secure door closed.
16. RESET the consumable reminder as described in
Section 5.3, Step 3.
9.3
Replacement of UV lamp LC158
The UV lamp should be replaced in the following circumstances:
•
When indicated by the consumable alarm or after a
maximum of one year.
•
Water purity in the installation starts to deteriorate.
1.
ENSURE process is OFF and ISOLATE power.
2.
OPEN front of unit by removing securing screws.
3.
LOCATE UV assembly on left-hand side of unit.
4.
PULL the electrical flex entering top of UV assembly
and withdraw lamp.
WARNING! IT IS RECOMMENDED THAT DURING THE
HANDLING OF THE LAMP CUT-RESISTANT
GLOVES BE WORN.
5.
HOLD lamp in gloved hand and PULL firmly on
electrical connector to remove.
6.
DISCARD lamp.
CAUTION!
Dispose of lamp in accordance with local
authority regulations.
7.
REMOVE new lamp from packaging and follow the
instructions included for cleaning.
8.
REFIT into UV assembly.
9.
RECONNECT to electrical connector ENSURING
correct orientation.
WARNING! LIGHT FROM THE UV LAMP IS EXTREMELY
HARMFUL TO THE EYES AND SKIN! UV
LAMPS SHOULD ONLY BE USED IN THE
REACTION CHAMBER WITH
SUITABLE
PROTECTION CAPS FITTED. PERSONS
SHOULD NEVER BE EXPOSED TO LIGHT
FROM THE LAMP.
10. REFIT rubber boot to stop UV light from being
radiated from top of the assembly.
11. RESET the consumable reminder as described in
Section 5.3, Step 4.