CENTRA – MDS/LDS/RDS US Operator Manual
ELGA
Page 36
CENTRA – MDS/LDS/RDS US Version 01 – 08/14
9.
MAINTENANCE
An approved supplier or distributor should carry out any
maintenance work not detailed in this handbook. If further
information is required on any aspect of maintenance please
contact Customer Service.
WARNING! ALWAYS CHECK THAT THE MAINS
ELECTRICAL POWER AND FEED WATER
SUPPLIES ARE SWITCHED OFF BEFORE
ATTEMPTING TO CHANGE A CONSUMABLE
ITEM.
9.1
Replacing the Composite Vent Filter / Air Filter
Air is drawn into the reservoir when water is used. To maintain
water purity it is necessary to filter the air. Air-borne particles and
bacteria are removed by an air-filter. Alternatively an improved
level of water quality will be maintained within the recirculation
system using a composite vent filter, which removes air-borne
contaminants such as volatile organic compounds and carbon
dioxide.
The Composite Vent Filter/Air Filter should be replaced in the
following circumstances:
•
Six months after initial installation.
•
Reservoir overflow - indicated by incorrect reservoir
volume.
•
Quality tests indicate replacement is required.
Step 1 - Replacement of CVF filter LC156/LC173
1.
ENSURE the process is OFF.
2.
OPEN the front of the unit by removing securing
screws.
3.
(LDS & RDS only) LOCATE the overflow device.
4.
SLIDE the overflow device off its location bracket
enabling access to the Bacterial Vent Filter.
5.
LOCATE the CVF.
6.
REMOVE the hose connection to the filter by
unscrewing the nut by hand and pulling the hose
towards you.
7.
REMOVE the filter from its location by PULLING
towards you.
8. DISCARD the filter.
9.
UNPACK the new filter.
10. INSTALL the filter into its location.
11. RECONNECT the hose.
12. WRITE the installation date on the label of the filter
for future reference.
13. RETURN the overflow device to its bracket.
CENTRA – MDS/MDS US (HF)
CVF filter location
CENTRA - LDS/RDS/RDS US (HF)
CVF filter location
CENTRA CVF filter connection
CVF Filter
CVF Filter