CENTRA – MDS/LDS/RDS US Operator Manual
ELGA
Page 16
CENTRA – MDS/LDS/RDS US Version 01 – 08/14
Step 2a - Electrical supply connection
(CENTRA – MDS)
WARNING! FAILURE TO CORRECTLY INSTALL THE
ELECTRICAL SUPPLY COULD RESULT IN
DEATH.
IF IN DOUBT CONTACT A QUALIFIED
ELECTRICAN.
1.
ENSURE a suitable cable is available to connect the
CENTRA
to the electrical supply.
WARNING! FAILURE TO USE THE CORRECTLY RATED
CABLE COULD RESULT IN FIRE AND DEATH.
IF IN DOUBT CONTACT A QUALIFIED
ELECTRICAN. CHECK LOCAL
REGULATIONS, A LICENCED ELECTRICIAN
MAY BE REQUIRED TO MAKE THIS
CONNECTION.
2.
REMOVE the four securing screws that hold the
front panel onto the unit.
3.
REMOVE the panel and store in a safe place.
4.
LOCATE the electrical cover on the inside of the unit
and REMOVE the four securing screws.
5.
CAREFULLY remove the cover and locate the
central earthing point on the electrical back plate.
6.
REMOVE the nut and washers securing the earthing
cable and store in a safe place.
Note: The electrical cover can now be completely
removed and stored safely.
7.
DIRECT the cable through the rear of the unit to the
power socket.
Note: The power cable should be routed avoiding
signal cables and sensitive circuitry.
8.
LOCATE the cable restraint on the left of the
electrical back plate.
9.
UNSCREW the cable clamp and INSERT the supply
cable ensuring sufficient cable to reach the MCB
and earth point.
10. RETIGHTEN the screws.
11. PULL the cable firmly and ENSURE that the cable is
properly retained.
12. CONNECT the supply cable to the MCB ensuring
correct termination and polarization.
CAUTION!
The termination of cables should comply with
local wiring regulations. All earth connections
should be made using ring tags that are
connected to the cable using the correct
crimping tool.
13. CONNECT the supply earth to the central earthing
point on the electrical back plate along with the earth
wire from the electrical cover.
CENTRA – MDS US rear view
Cable entry