Velp Scientifica F102B0153 Instruction Manual Download Page 7

3.5 Scale selection 

Scale 

Volume of sample

 

0 ÷ 90 mg O

2

/l 

400 ml 

0 ÷ 250 mg O

2

/l 

250 ml 

0 ÷ 600 mg O

2

/l 

150 ml 

0 ÷ 1000 mg O

2

/l 

100 ml 

0 ÷ 4000 mg O

2

/l 

25ml 

PRE 

Refer to the method 

 
The scale is related to the expected value. The possibility of measuring diluted samples extends the range

Results are shown as mg O

2

/l (ppm) for BOD test and values in mbar for PRE scale. 

 
Wireless mode  

 

 

 

 

 

Non-wireless mode (after the reset) 

Before the analysis set the scale on the RESPIRO

Soft™ 

Key 

SET

 (display on) 

 

The scale is changed 

Key 

SET

 or 

START

  

 

The scale is shown 

3.6 Analysis duration and sampling time 
 
Wireless mode  

With RESPIROMETRIC Sensor an analysis can last from 5 to 
120 days. The sampling time can be selected depending. 
on the desired duration of the analysis, from 1 h to 2 days 

 
Wireless mode  

Before the analysis set the scale on the RESPIROSoft

™ 

3.7 Alarms 
AL1

 

Calibration error. Restart the analysis. If the problem persists contact the service. 

AL2

 

Analysis start denied. Check on RESPIROSoft

™. 

AL3

 

Max. sensors q.ty reached. 50 analyses are already in progress. Wait the end. 

AL5

 

No PC response. Check the distance between the  RESPIROMETRIC Sensor 

and the Wireless DataBox™. 

AL8

 

Data sending failed. Restart the analysis. If the problem persists contact the service. 

AL9/11

 No PC response. Check if the RESPIROSoft

™ is active and if the Wireless DataBox™ connected. 

bAt

 

Low  battery  (wireless  mode  OFF).  Remove  the  batteries  with  the  fingers  and  replace  them  with  new  ones  paying 
attention to the polarity indicated inside the instrument. 

Err

 

Calibration error (wireless mode OFF). 

4.  Maintenance 

RESPIROMETRIC Sensor warns about the required change of  batteries when the unit is resetted and the charge is lower 
than a minimum threshold. Use a coin lever in the rear part of RESPIROMETRIC Sensor and open it. Batteries are removed 
using the fingers. Introduce two fresh batteries at the same time paying attention to the polarity. Battery replacement deletes 
any stored value. 

5.  Technical data 

Dimensions (WxHxD) 

50x70x70 mm 

Weight 

80 g 

Measurement 

Electronic pressure probe 

BOD values or pressure 

Directly on the 
RESPIROSoft

™,  

at any time 

Stored data 

5 BOD values per sensor; 
Database on RESPIROSoft

™ 

Display 

4 LED digits, 7 mm high 

Precision 

± 1 digit (

 ±3.55 hPa) ± 1% 

Operation range 

500 ÷ 2000 mbar (h Pa) 

Scales 

90, 250, 600, 999 4000 ppm 
BMP PRE  

Power supply 

2 lithium batteries (250 mAh) 

Electrical input with display on  15 mA max. 

Electrical input with display off  7,5 µA 

Protection degree 

IP 54-EN 60529 

Admitted room temperature 

Storing 

–25 

 +65° C               

Ope5 

 +40°C 

6.  Accessories / Spare parts 

A00000135 

BOD Sensor Check 

 

 

A00000136 

Control Test Tablets  

10000904 

Alkali container 

 
 
 

Non-wireless mode

  

Respirometric Sensor will perform only 5-day test 
with 24-hour sampling time for BOD analysis or 
30 days using PRE scale.

 

Summary of Contents for F102B0153

Page 1: ...sto strumento deve essere utilizzato solo per applicazioni di laboratorio La societ produttrice declina ogni responsabilit sull impiego non conforme alle istruzioni degli strumenti Cet instrument ne p...

Page 2: ...rmit tserkl rung We the manufacturer VELP Scientifica under our responsibility declare that the product is manufactured in conformity with the following standards Noi casa costruttrice VELP SCIENTIFIC...

Page 3: ...STAZIONE DELLA SCALA 9 3 6 DURATA ANALISI E TEMPO CAMPIONAMENTO 9 3 7 ALLARMI 9 4 MANUTENZIONE 9 5 CARATTERISTICHE TECNICHE 9 6 ACCESSORI PARTI DI RICAMBIO 9 1 INTRODUCTION 10 2 ASSEMBLAGE ET INSTALLA...

Page 4: ...MODUS 14 3 3 BESCHREIBUNG DER TASTEN VERKABELTER MODUS 14 3 4 WIEDERHERSTELLUNG DER LETZTEN DATEN 15 3 5 WAHL DES MESSBEREICHS 15 3 6 DAUER DER ANALYSEDAUER UND ZEIT DER PROBENAHME 15 3 7 ALARME 15 4...

Page 5: ......

Page 6: ...3 2 Description of keys wireless mode Out of a measuring cycle before starting Key SET with display off Wireless symbol is shown Key SET with display on Wireless symbol is shown Key START with displa...

Page 7: ...IROSoft is active and if the Wireless DataBox connected bAt Low battery wireless mode OFF Remove the batteries with the fingers and replace them with new ones paying attention to the polarity indicate...

Page 8: ...3 2 Descrizione dei tasti in modalit wireless Fuori dal ciclo di misure prima dello Start Tasto SET a display spento Visualizza simbolo wireless Tasto SET a display acceso Visualizza simbolo wireless...

Page 9: ...5 Nessuna notifica PC Verificare che la distanza tra sensore e PC sia idonea AL8 Invio dati fallito Far ripartire l analisi In caso il problema persista contattare il servizio di assistenza AL9 11 Nes...

Page 10: ...touches mode sans fil En dehors d un cycle de mesure avant le d marrage Touche SET avec cran off Le symbole sans fil est affich Touche SET avec cran on Le symbole sans fil est affich Touche START avec...

Page 11: ...le service AL9 11 Pas de r ponse du PC V rifier si RESPIROSoft est activ et si la DataBoxTM sans fil est connect e bAt Batterie faible mode sans fil OFF Enlever les batteries avec les doigts et les r...

Page 12: ...Descripci n de las teclas modo inal mbrico Fuera de un ciclo de medici n antes de comenzar Tecla SET con la pantalla apagada Se muestra el s mbolo de modo inal mbrico Tecla SET con la pantalla encend...

Page 13: ...en el env o de datos Reinicie el an lisis Si el problema contin a p ngase en contacto con el servicio de atenci n al cliente AL9 11 No hay respuesta del PC Compruebe que el RESPIROSoft est activo y qu...

Page 14: ...s ON Taste START bergang zum kabellosen Modus OFF 3 2 Beschreibung der Tasten kabelloser Modus Au erhalb eines Messzyklus vor dem Start Taste SET mit ausgeschaltetem Display Das Kabellos Symbol wird a...

Page 15: ...rollieren AL8 Senden der Daten fehlgeschlagen Die Analyse neu starten Wenn das Problem weiterhin besteht den Kundendienst kontaktieren AL9 11 Keine Antwort vom PC Pr fen ob der RESPIROSoft aktiviert u...

Page 16: ...20 C 0 5 BMP BOD 2 RESPIROMETRIC RESPIROMETRIC 2 3 3 RESPIROMETRIC 3 1 RESPIROMETRIC SET S E T SET START 3 2 SET SET START START SET START 5 START 5 3 3 SET SET START START SET START 5 START 5 SET ST...

Page 17: ...0 1 2 RESPIROSoft 3 7 AL1 VELP AL2 RESPIROSoft AL3 q ty 50 AL5 PC RESPIROMETRIC Wireless DataBox AL8 AL9 11 PC RESPIROSoft Wireless DataBox bAt Err 4 RESPIROMETRIC RESPIROMETRIC 5 WxHxD 50x70x70 80 BO...

Page 18: ...18 6 A00000135 BOD A00000136 10000904...

Page 19: ......

Page 20: ...zers Digestion Units Distillation Units Solvent Extractors Fiber Analyzers Dietary Fiber Analyzers Oxidation Stability Reactor Consumables Analytical Instruments Analizzatori Elementari Digestori e Mi...

Reviews: