14
4 Mounting
VEGAPULS Air 23 • Autarkic device with measured value transmission via radio technology
64314-EN-230606
1. Loop the tension belt into the cross struts of the IBC container
2. Insert belt into guides on the housing cover, close click connec-
tion
2
3
1
Fig. 7: Mounting flexible exchangeable bracket
1 Vessel top
2 Foam pad
3 Tension belt
3. Slide sensor into mounting position
4. Tighten belt, check sensor for secure fit
Note:
For optimum measurement, ensure direct, permanent contact be-
tween the bottom of the sensor and the surface of the IBC container.
The device version for ceiling mounting has mounting brackets on the
housing cover. The mounting is carried out using suitable screws and
dowels provided by the customer.
Fig. 8: Ceiling mounting with screws
The device version for ceiling mounting with magnets has three
magnets on the housing lid. This allows the sensor to be attached to
ferromagnetic surfaces without drilling or screwing.
Fig. 9: Ceiling mounting with magnets
Ceiling mounting with
screws
Ceiling mounting with
magnets