45
13 Supplement
VEGAPULS 42 • Three-wire with IO-Link (2 x transistor or 4 … 20 mA plus 1 x transistor)
1016153-EN-230316
Max. power consumption
9)
Ʋ
Sensor
3.5 W
Ʋ
Load per transistor output
10)
9 W
Reverse voltage protection
Integrated
Permissible residual ripple
Ʋ
for U
N
12 V DC (12 V < U
B
< 18 V)
≤ 0.7 V
eff
(16 … 400 Hz)
Ʋ
for U
N
24 V DC (18 V < U
B
< 35 V)
≤ 1 V
eff
(16 … 400 Hz)
Max. load resistor
Ʋ
Operating voltage U
B
= 12 V DC
370 Ω
Ʋ
Operating voltage U
B
= 18 V DC
630 Ω
Electrical protective measures
Potential separation
Electronics potential free up to 500 V AC
Protection rating
Connection technology
Protection according to
EN 60529/IEC 529
Protection according to NEMA/
UL 50E
M12 x 1 plug
IP66/IP67/IP69
Type 6P
Altitude above sea level
≤ 5000 m (≤ 16404 ft)
Protection class
III
Pollution degree
4
13.2 IO-Link
In the following, the necessary device-specific details are shown. You can find further information of
Physical layer
IO-Link specification: Revision 1.1
SIO mode: Yes
Speed: COM2 38.4 kBaud
Min. cycle time 5.0 ms
Length process data word: 48 Bit
IO-Link Data Storage: Yes
Block parameter adjustment: Yes
Direct parameter
Byte
Parameter
HexCode
Remark, value
0
-
-
-
2
MasterCycleTime
-
-
3
MinCycleTime
0x28
5 ms
9)
U
B
= 35 V DC, output signal = 20 mA
10)
Load current = 250 mA