P
roceed as follows
:
1
E
dit the mbar value in the menu item
"
zero
"
with
[
OK
]
.
Z
ero adjustment
000
.
0
%
P
=
+
0000
.
0
mbar
0000
.
0
mbar
2
S
et the requested mbar value with
[
+
]
and
[–
>
]
.
3
C
on
fi
rm with
[
OK
]
and move to span adjustment with
[–
>
]
.
T
he zero adjustment is
fi
nished
.
I
nformation
:
T
he zero adjustment shifts the value of the span adjustment
.
T
he span
,
i
.
e
.
the di
ff
erence between these values
,
however
,
remains unchanged
.
I
nformation
:
T
o adjust with pressure
,
you simply enter the indicated actual
measured value
.
I
f the adjustment ranges are exceeded
,
the
message is displayed
"
O
utside parameter limits
"
.
T
he editing
procedure can be interrupted with
[
ESC
]
or the displayed limit
value can be accepted with
[
OK
]
.
P
roceed as follows
:
1
E
dit the mbar value in the menu item
"
span
"
with
[
OK
]
.
S
pan adjustment
100
.
0
%
P
=
+
1000
.
0
mbar
0000
.
0
mbar
I
nformation
:
T
he displayed pressure for
100
%
corresponds e
.
g
.
to the
nominal range of the sensor
(
in the above example
1
.
0
bar
=
1000
mbar
).
2
S
et the requested mbar value with
[–
>
]
and
[
OK
]
.
3
C
on
fi
rm with
[
OK
]
and move to the menu overview with
[
ESC
]
.
T
he span adjustment is
fi
nished
.
C
arry out zero adjustment
C
arry out span adjustment
VEGABAR
53
-
P
ro
fi
bus
PA
25
S
etup with the indicating and adjustment module
PLICSCOM
28250
-
EN
-