Pax-i3D Smart(PHT-30LFO) Installation Manual
23
2 Choosing an Installation Site
2.2
Specifications for Electrical Installation
These specifications are based on the
MEIGaN
(Medical electrical installation guidance notes).
Consult the companion manual for further information. :
Volume 3: Specification for Electrical
Installation.
2.3
Electrical Requirements
This equipment must be connected to a grounded outlet to fulfill the safety
provisions specified in
IEC 60364: the 2
nd
edition (2006).
It is mandatory that both PC and equipment use the same power line if
connected to an MPSO.
Whenever possible, use different power outlets for each device. If a
multiple portable socket outlet (MPSO) must be used, ensure that the PC
and equipment are connected to the same MPSO.
Use a dedicated power outlet for the power cord. Failure to do so may
result in unstable system operation caused by power fluctuations.
It is strongly recommended that you install an AVR. An AVR (automatic
voltage regulator) maintains a constant voltage and allows for continuous
operation in the event of a power fluctuation.
Power supply voltage
AC100-240 V
●
The input line voltage depends on the local electrical distribution system.
● Allowable input voltage fluctuation requirement: ± 10 %.
Frequency
50/60 Hz
Phase
single
Power rating (maximum power consumption)
Max.2.2 kVA (during exposure)
Summary of Contents for PaX-i3D Smart
Page 1: ...PHT 30LFO HARDWARE and SOFTWARE E n g l i s h Installation Manual Version 1 0 0...
Page 2: ......
Page 3: ......
Page 72: ...This page is intentionally left blank...
Page 82: ...This page is intentionally left blank...
Page 92: ...This page is intentionally left blank...
Page 137: ...10 Technical Specifications...
Page 145: ...Pax i3D Smart PHT 30LFO Installation Manual 145 Appendix Schematic diagram Components supplied...
Page 160: ......
Page 161: ...13 Samsung 1 ro 2 gil Hwaseong si Gyeonggi do Korea Postal Code 445 170...