A.3 Electromagnetic
Compatibility (EMC) Information
Guidance and manufacturer’s declaration - electromagnetic emissions
The VEX-S300W is intended for use in the electromagnetic environment specified
below. The customer or the user of the VEX-S300W should assure that it is used in
such an environment.
Emissions test
Compliance
Electromagnetic
environment - guidance
RF emissions CISPR 11
Group 1
The VEX-S300W uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low and
are not likely to cause any
interference in nearby
electronic equipment.
RF emissions CISPR 11
Class A
The VEX-S300W is suitable
for use in all establishments
and may be used in domestic
establishments and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Applicable
Voltage fluctuations / flicker
emissions
IEC 61000-3-3
Applicable
NOTE) It is essential that the actual RF shielding effectiveness and filter attenuation
of the shielded location be verified to ensure that they meet or exceed the specified
minimum values.
Summary of Contents for EzRay Air W
Page 1: ......
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...Notice vi EzRay Air W Model VEX S300W User Manual This page intentionally left blank...
Page 56: ...8 Product Specifications 48 EzRay Air W Model VEX S300W User Manual...
Page 57: ...8 Product Specifications EzRay Air W Model VEX S300W User Manual 49 ENGLISH...
Page 84: ......
Page 85: ......
Page 86: ......