A.2 X-ray Dose Data
The X-ray dose data is extracted from the X-ray Dose Test Report for the VEX-S300W.
The X-ray doses of VEX-S300W in the test report have been measured in accordance
with the IEC collateral standards. The VEX-S300W has been designed in accordance
with Part 1. General Requirements for Safety, IEC 60601-1-3.
Test Condition
Model Name
VEX-S300W
Tube Model Name
V1-650304
Generator Model Name
DG-S0102V1 (Inverter type)
Loading Factor
Option A: 65 kV, 2.5 mA
Option B: 65 kV, 3.0 mA
1. X-ray Dose Table
Test Equipment
Instrument
Manufacturer
Model
S/N
Multi Dose Meter
UNFORS
Unfors Xi
R/F&MAM
161834
Dose Table (65 kVp, 2.5 mA, FOV: Ø 6 cm, SSD 200 mm, at Al 6 mm)
t (s)
Dose (
μ
Gy)
0.20
107
0.22
118
0.24
128
0.25
134
Dose Area Product (DAP) Table (65 kVp, 2.5 mA, SSD 200 mm)
FOV: Ø 6 cm
FOV : 4 x 3 cm
FOV : 2 x 3 cm
t (s)
Dose (mGy.cm
2
)
0.05
2.19
0.93
0.47
0.08
3.92
1.67
0.83
0.10
5.08
2.16
1.08
0.15
7.96
3.38
1.69
0.20
10.84
4.30
2.60
0.25
13.73
5.83
2.91
Summary of Contents for EzRay Air W
Page 1: ......
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...Notice vi EzRay Air W Model VEX S300W User Manual This page intentionally left blank...
Page 56: ...8 Product Specifications 48 EzRay Air W Model VEX S300W User Manual...
Page 57: ...8 Product Specifications EzRay Air W Model VEX S300W User Manual 49 ENGLISH...
Page 84: ......
Page 85: ......
Page 86: ......