6. User Maintenance
EzRay Air W (Model: VEX-S300W) User Manual
43
EN
G
LISH
6. User Maintenance
To ensure that the equipment operates at maximum efficiency, it is recommended to
perform the following procedures for the customer’s routine maintenance activities.
Maintenance Task Checklist
Always turn off the equipment before performing any maintenance.
Tasks
Period
Sterilize all components that come into contact with the patient
and operator by using an alcohol-based solution.
Daily
Wipe the outer covers of the equipment with a dry cloth at the
end of each day’s operation.
Daily
Ensure that the main power button has been turned off after
using the equipment.
Daily
Ensure the audible signal is audible and the X-ray exposure
light is visible when you perform an exposure.
Daily
Ensure that the wall framework is securely attached to the wall.
Daily
Ensure that the yellow (exposure) indicator light turns on when
the exposure button is pressed.
Daily
Ensure that the power cable does not have cuts or abrasions.
Monthly
Ensure that all cable connectors are not mechanically
defective.
Monthly
Ensure that the control panel has no defects.
Monthly
Ensure that the power cable does not have cuts or abrasions.
Monthly
Ensure that all visible labels are intact and legible.
Monthly
DO NOT use detergents or solvents to clean the outer covers of the
equipment.
If any defects are found, do not operate the equipment since it has to
be handled by a qualified person. Contact your Service
Representative.
Summary of Contents for EzRay Air W
Page 1: ......
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...Notice vi EzRay Air W Model VEX S300W User Manual This page intentionally left blank...
Page 56: ...8 Product Specifications 48 EzRay Air W Model VEX S300W User Manual...
Page 57: ...8 Product Specifications EzRay Air W Model VEX S300W User Manual 49 ENGLISH...
Page 84: ......
Page 85: ......
Page 86: ......