5. Troubleshooting
EzRay Air W (Model: VEX-S300W) User Manual
41
EN
G
LISH
5. Troubleshooting
In instances of abnormal operation, error messages will be displayed on the Control
Panel. If a problem persists, please request assistance from the customer support
information services.
Error Messages
Error Code
Check Parameter
Description
E.02
X-ray Generator
The kV feedback is lower than the
set value during X-ray irradiation.
E.03
X-ray Generator
The kV feedback is higher than the
set value during X-ray irradiation.
E.04
X-ray Generator
The mA feedback is lower than the
set value during X-ray irradiation.
E.05
X-ray Generator
The mA feedback is higher than the
set value during X-ray irradiation.
E.06
X-ray Generator
The mono-tank block temperature
exceeds the error limit.
E.07
System
The Remote Exposure Switch was
pressed when the system power is
turned on.
E.08
User
1) The Remote Exposure Switch was
turned off earlier than the exposure
time.
2) The door of the X-ray room was
opened during X-ray exposure when
the door has the Door Interlock
Switch (option) installed.