5. Troubleshooting
42
EzRay Air W (Model: VEX-S300W) User Manual
Troubleshooting
Problem
Cause
Solution
Equipment is not
turned on.
Power button is not turned
on properly.
Turn the equipment power
switch off and turn it back on.
Control Panel is not
turned on.
Defective main board
Contact your Service
Representative.
Internal cable disconnected Contact your Service
Representative.
No X-ray emission
Generator is cooling.
Wait for the cooling time. (refer
to ‘Duty Cycle’ on
page 51.
)
Remote Exposure Switch is
pressed, but X-ray
exposure does not work.
Contact your Service
Representative.
Internal cable disconnected Contact your Service
Representative.
Defective generator
Contact your Service
Representative.
Tube lifecycle termination
Contact your Service
Representative.
X-ray emission
works, but
exposure is too light
or
completely white.
Equipment has been
positioned incorrectly.
Adjust the position of the
equipment.
Exposure time is too short.
Increase the exposure time.
Receptor is facing the
wrong way.
Reposition the receptor.
X-ray emission
works, but
exposure is too
dark.
Exposure time is too long.
Decrease the exposure time.
Summary of Contents for EzRay Air W
Page 1: ......
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...Notice vi EzRay Air W Model VEX S300W User Manual This page intentionally left blank...
Page 56: ...8 Product Specifications 48 EzRay Air W Model VEX S300W User Manual...
Page 57: ...8 Product Specifications EzRay Air W Model VEX S300W User Manual 49 ENGLISH...
Page 84: ......
Page 85: ......
Page 86: ......