17. Remove the exhaust fitting.
18. Remove and discard the O-ring.
19. Using the hemostat pliers, remove the large
spring, plunger guide, small spring and
check valve from the exhaust port in the
inboard housing.
TriScroll 300 Disassembly
(continued)
NOTE
The next step applies to all
models except the 310.