Introduction and Initial Operation
Page 10
User Manual Micro-GC
Varian, Inc.
INSTALLATION
Install Gas Regulators and Set pressures
Carrier gas supplied from cylinders should have a two-stage pressure assembly to
adjust the carrier gas pressure to 550 ± 10 % (80 ± 10 %). Set cylinder regulator
pressure to match the Micro-GC gas inlet pressure
(see
).
Carrier gas
The most commonly used carrier gas for the Micro-GC is either He or N
2
. The
recommended purity for carrier gas is 99.995% minimum.
See
Connect data handling peripherals
The Micro-GC must be connected to an
both to set up
the method and to acquire data.
Workstations and other data handling devices are connected to the cable connectors
inside the Micro-GC (see
Hardware connection between the Micro-GC and Workstation (computer) is done
by Ethernet connection. For additional information look here
Sample
Connect the sample to the using the sample-in connector situated at the front of the
instrument (
Read
for more important information!
Heated Sample line
To connect heated sample lines refer
.