Introduction and Initial Operation
Varian, Inc.
User Manual Micro-GC
Page 27
COLUMN HAYESEP 40CM HEATED
The HaySep A column separates oxygen, methane, carbon dioxide, ethane, acetylene,
ethylene, and selected sulfur gases. Nitrogen coelutes with oxygen. Components with a
higher molecular weight than propane have long retention times on this column.
Maximum allowable column temperature: 160
C
Instrument parameters
Column temperature
: 50°C
Injector temperature
: 110°C
Column pressure
: 150kPa (21PSI)
Sample time
: 30s
Injection time
: 40ms
Run
time
:
60s
Detector sensitivity
: Auto
Peak identification
ID Component Concentration
1 Nitrogen
0.77%
2 Methane
Balance
3 Ethane
8.1%