COLUMN CP-SIL 5 CB 6 METER UNHEATED
The natural gas components, mostly hydrocarbons, separate in the same order on the
non-polar and medium-polar CP-Sil CB columns. Nitrogen, methane, carbon dioxide,
and ethane are not separated on these columns. They produce a composite peak. For
separation of these components a HayeSep A column is advised.
Instrument parameters
Column temperature
: 50°C
Injector temperature
: NA
Column pressure
: 150kPa (21PSI)
Sample time
: 30s
Injection time
: 40ms
Run
time
:
60s
Detector sensitivity
: Auto
Peak identification
ID Component Concentration
1 Composite Balance
2 Ethane
8.1%
3 Propane
1.0%
4 i-Butane
0.14%
5
n-Butane
0.20%