CP-3800 GC Getting Started Manual
9
Installation
Site Requirements for GC Installation
Power
Requirements
Power Source
Each 3800 GC requires a separate circuit with the following
characteristics:
101 Vac
±
10%, 50 or 60 Hz
±
2%, 25 Amps, 2.5 Kilowatts
120 Vac
±
10%, 60 Hz
±
2%, 20 Amps, 2.4 Kilowatts
230 Vac
±
15%, 50 Hz
±
2%, 10 Amps, 2.3 Kilowatts
Installation Category: II
Ground
It is essential to have a safety ground connection to this instrument.
Power Cord
In Europe the instrument will be supplied with an IEC 320 style socket.
The power cord assembly can be changed to suit your local power and
socket requirements. In the United States and Japan the instrument is
equipped with a fixed power cord and plug.
Heat Dissipation
Calculate heat dissipation using the following formula:
Watts x 3.4129 = BTU/hr.
Actual dissipation will depend on method parameters and instrument
configuration.
C A U T IO N
Heat from the column oven is vented through the rear of the GC. Make
certain that the hot column oven exhaust is not directed onto other
electronic instruments, cables or gas lines.
Fuses
There are four fuses accessible on the top of the 3800:
F1: 6.3A T, F2: 6.3A T, F3: 10.0A F, F4: 6.3A T