1061 Injector
The 1061 is mechanically similar to the 1041 except that it uses a glass insert
for flash vaporization injection. Follow the same maintenance procedures for
the 1061 that are outlined above for the 1041, except in the case of changing
or cleaning glass inserts.
Replacing
the 1061
Glass Insert
After prolonged use, the 1061 glass insert may need to be removed for
cleaning purposes or to be replaced with a new insert. Note that there are two
different glass inserts, one for use with large-bore (0.53 mm ID) capillary
columns and one for use with packed columns. These inserts are not
interchangeable.
•
Large-bore glass insert, Part Number: 03-918339-00
•
Packed column glass insert, Part Number: 37-000813-00
Refer to the figure below when replacing the glass insert.
Figure 5 1061 Injector Assembly
1
Septum Nut
2
Automatic Start Switch
3
Septum, 10 mm
4
Injector Insert
5
Injector Body
These tools are needed to replace the glass insert:
•
Tweezers or septum pick (P/N 72-000084-00)
•
Injector nut wrench (P/N 03-908423-00)
•
Clean laboratory tissue
1
2
3
4
5