Technical Manual: TM80030
THERMICA WATER TO AIR COOLER
Model range: THERMICA10-VV-MMM-LLL
Revision: 1.0
Date of release: 2022-12-16
15/55
www.vareximaging.com
2.
Device description
This chapter will describe the device. The device is pointed out with features present in §2.4.
Classifications of the device are listed in §2.7.
2.1.
INTENDED USE AND USERS
The device is intended to be used for the cooling of X-ray and E-beam tubes in industrial applications.
The device is intended to be used in stationary applications in a fixed position and industrial environment
like:
-
Non-destructive testing applications
-
Laboratory and research environments
-
Therapeutic x-ray equipment for radiotherapy treatments or sterilization packaging material by
irradiation
-
Industrial machine cooling
The limitations of use are specified in §9.
The intended user of the device is defined as the operator of an X-ray or E-beam system. The operator
shall be qualified and/or trained to operate the X-ray or E-beam system.
2.2.
Description of the device
The THERMICA WATER TO AIR COOLER is operated by the generator and consists of a minimum flow
switch, level switch and a temperature controller which will generate a monitoring signal to the generator
once the minimum system requirements are not met.
In addition, the device can be configurated with an Air filter.
The device provides 2 MOP:
-
between the electronics and the outside of the housing
-
between the electronics and the operator controls
-
between motor and fan and other components and encasing
The device is suitable for installation in stationary X-ray systems.
For configuration and usage see §5 and §6.