Technical Manual: TM80030
THERMICA WATER TO AIR COOLER
Model range: THERMICA10-VV-MMM-LLL
Revision: 1.0
Date of release: 2022-12-16
28/55
www.vareximaging.com
4.6.
Wire harness
There are six wire harnesses present in the cooler. Each wire harness has its own Part Number printed
on a label. The wire harnesses can be ordered as service parts. (see §9.6)
4.7.
EMC compatibility
The device conforms to IEC61326-1:2021 for EMC compatibility and must be installed and put into
service according to the EMC information provided in this manual.
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the
Cooler, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
WARNING:
Not taking EMC measures into account on the wiring may result in increased EMISSIONS or
decreased IMMUNITY. IEC61326-1:2021 must be followed for being complaint with EMC
guidelines.
WARNING:
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are operating normally.
WARNING:
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
NOTE:
The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas. The cooler is not designed for use in a residential environment (for which CISPR 11
class B is normally required). This equipment might not offer adequate protection to radio-
frequency communication services in residential environment.
Deviations
No deviations from IEC61326-1:2021 are applied.
Allowances