background image

V&P Scientific, Inc. 

Technical Note 396C 

 

  

 

 
 

OPERATION 

 

POWERING UP THE

 

SpinVessel® SYSTEM 

The Power Switch (I/O) is located on the back of the Controller. To power up the C

ontroller, depress the “I” 

of the switch.   

 

OPERATING THE

 

SpinVessel® With COMPUTER CONTROL 

The SpinVessel® Models with “

-

CC” in the part number are contr

olled using a computer.  The Controller 

for the computer controllable SpinVessel® is connected by a USB communication cable to a computer. 
The Controller accepts simple ASCII commands to set the speed of the rotations, the number of rotations 
(including fractions of a rotation), duration of pause before reversing direction and degree of ramping to 
speed set. 

 
Software details 

1.

 

The SpinVessel® Controller uses an internal CH340G USB to Serial Converter (set to 9600,8,N,1) 
which will require driver software.  Windows may already have the needed USB driver software 
installed.  If not, download 

the “CH340 USB DRIVER” 

from the provided USB flash drive and run 

“CH341SER.exe” to install it (Windows typically auto

-installs once downloaded).  If your company 

policy does not allow the use of the provided USB stick, this driver is also available for download 
from the internet (contact your IT department for assistance if necessary). 

2.

 

The basic command language for this computer-to-stirrer communication is ASCII. 

3.

 

Commands can be sent from the computer using a standard terminal program like HyperTerminal 
or Realterm.  Or, if needed, download the V&P Scientific Serial Terminal program from the 

With the SpinVessel® 
tube in the drive base, 
check the tip alignment 
relative to the projections. 
 

Without liquid in the tube, 
turn on the SpinVessel® to 
make sure the tips are not 
hitting the projections. 

Summary of Contents for SpinVessel VP 418SV2-1-300FB-CC

Page 1: ...11 623 188 Foreign Patents Pending SpinVessel System with Computer Control feature for 300ml SpinVessel Tube VP 418SV2 1 300FB CC includes Motor Unit Spin Base Controller Power Supply not shown Registration Disc not shown USB cable and USB drive with control software not shown Shown with SpinVessel Tube VP 830SV 300FB sold separately ...

Page 2: ... SpinVessel SYSTEM Power and Controller Connections Power Switch Power Cable Receptacle USB Cable Receptacle Controller Cable Receptacle Controller Cable Controller Motor Unit Power Supply Power Cable Power Cord USB Cable Spin Base Parts of the System Controller USB Drive ...

Page 3: ...er Supply to Controller Warning Do not plug Controller in while power is on Ensure that the Power Switch is in the off position 0 is down Always have Controller plugged in before powering up 5 Connect Controller Cable from the Motor Unit to the Controller 6 Connect the USB Cable from the Controller to the computer 7 Place one VP 830SV 650CFB SpinVessel tube into the Spin Cup Ensure that the SpinVe...

Page 4: ...ate within drive base Place SpinVessel motor unit in desired location either on an automated liquid handler or on a benchtop The motor unit is SBS SLAS format 127 7mm x 85 5mm which will fit into any SBS SLAS plate locator or carrier To align pipet tips to inside of SpinVessel tube use the registration disc VP 418SV 300FB Place disc on top of the drive base not the tube Lower the tips until they j...

Page 5: ...Serial Converter set to 9600 8 N 1 which will require driver software Windows may already have the needed USB driver software installed If not download the CH340 USB DRIVER from the provided USB flash drive and run CH341SER exe to install it Windows typically auto installs once downloaded If your company policy does not allow the use of the provided USB stick this driver is also available for down...

Page 6: ...am V P Scientific version or other 3 Turn on the SpinVessel instrument as described 4 Select the appropriate COM port and click CONNECT to establish a line of communication between the instrument and the computer 5 Create ASCII command strings as outlined in the diagram shown below to control the instrument with the following parameters a Speed of rotation in revolutions per minute RPM b Number of...

Page 7: ...n The following are settings to use with an aqueous solution when first testing the VP 418SV 300FB CC SpinVessel Speed 50 RPM Rotate 2 1 Pause 200 milliseconds Ramp 10 It is always a good idea to practice with water in your SpinVessel to determine safe RPM Rotation Pause and Ramp numbers before you use expensive reagents Note that if Ramp is set to 0 liquid will splash out of the Spin Vessel The V...

Page 8: ...e to comply with safety requirements and practices WARNING To avoid personnel injury caused by electrical shock do not remove the cover of the controller when the power is ON CAUTION Do not disconnect motor during operation Otherwise over current breakdown may result WARRANTY V P Scientific Inc warrants this product to be free from defects in material and workmanship when used under normal laborat...