V&P Scientific, Inc. 

Technical Note 396C 

USING A 300 ML SpinVessel

®

 SYSTEM 

The optimal stirring mode is dependent upon the application and needs to be empirically 
determined. 

Factors to consider in determining optimal 300 mL SpinVessel® operation are the density 

of the particulates, their size and shape, as well as the volume and viscosity of the liquid.  Start with 50 
RPM and 2.1 rotations before reversing direction, then work up to the appropriate RPMs and number of 
rotations for the 

particulate’s density, fragility, and solution viscosity.  We have found that smaller 

circumference SpinVessels® require higher RPMs than larger circumference SpinVessels do, because the 
critical factor is the linear speed generated at the circumference not the RPMs.  See page 501 of our 

SLAS 

Technology paper

 for a discussion of this phenomenon. 

 
The  following are  settings  to  use  with  an  aqueous  solution  when  first testing  the  VP  418SV-300FB-CC 
SpinVessel®: 

Speed: 50 RPM 

Rotate: 2.1  Pause: 200 milliseconds 

     Ramp: 10  

 

It is always a good idea to practice with water in your SpinVessel to determine safe RPM, Rotation, Pause 
and Ramp numbers before you use expensive reagents.  
 

Note that, if Ramp is set to 0, liquid will splash out of the Spin Vessel.  
 

The VP 418SV-1-300FB-CC SpinVessel® is designed to be used only with a

 

VP 830SV-300FB SpinVessels® 

tube.   Please contact V&P Scientific for pricing. 
 
See below for an example of low versus high RPM for a solution of yeast cells in a large circumference 6 
liter SpinVessel®.  Note: That when given a longer mixing time even the slower speed (20 RPM i.e., 

 more 

gentle mixing) was still able to suspend the yeast cells. 

 

 

0.00

0.20

0.40

0.60

0.80

1.00

1.20

2

5

9

14 18 24 28 33 37 40 44 60 90 120 230 480 720

A

bso

rb

anc

e

Time (Seconds)

Yeast Resuspension with 6L SpinVessel®

20 RPM vs 100 RPM

20 RPM, 4 rotations, 500 pause

100 RPM, 4 rotations, 500 pause

Summary of Contents for SpinVessel VP 418SV2-1-300FB-CC

Page 1: ...11 623 188 Foreign Patents Pending SpinVessel System with Computer Control feature for 300ml SpinVessel Tube VP 418SV2 1 300FB CC includes Motor Unit Spin Base Controller Power Supply not shown Registration Disc not shown USB cable and USB drive with control software not shown Shown with SpinVessel Tube VP 830SV 300FB sold separately ...

Page 2: ... SpinVessel SYSTEM Power and Controller Connections Power Switch Power Cable Receptacle USB Cable Receptacle Controller Cable Receptacle Controller Cable Controller Motor Unit Power Supply Power Cable Power Cord USB Cable Spin Base Parts of the System Controller USB Drive ...

Page 3: ...er Supply to Controller Warning Do not plug Controller in while power is on Ensure that the Power Switch is in the off position 0 is down Always have Controller plugged in before powering up 5 Connect Controller Cable from the Motor Unit to the Controller 6 Connect the USB Cable from the Controller to the computer 7 Place one VP 830SV 650CFB SpinVessel tube into the Spin Cup Ensure that the SpinVe...

Page 4: ...ate within drive base Place SpinVessel motor unit in desired location either on an automated liquid handler or on a benchtop The motor unit is SBS SLAS format 127 7mm x 85 5mm which will fit into any SBS SLAS plate locator or carrier To align pipet tips to inside of SpinVessel tube use the registration disc VP 418SV 300FB Place disc on top of the drive base not the tube Lower the tips until they j...

Page 5: ...Serial Converter set to 9600 8 N 1 which will require driver software Windows may already have the needed USB driver software installed If not download the CH340 USB DRIVER from the provided USB flash drive and run CH341SER exe to install it Windows typically auto installs once downloaded If your company policy does not allow the use of the provided USB stick this driver is also available for down...

Page 6: ...am V P Scientific version or other 3 Turn on the SpinVessel instrument as described 4 Select the appropriate COM port and click CONNECT to establish a line of communication between the instrument and the computer 5 Create ASCII command strings as outlined in the diagram shown below to control the instrument with the following parameters a Speed of rotation in revolutions per minute RPM b Number of...

Page 7: ...n The following are settings to use with an aqueous solution when first testing the VP 418SV 300FB CC SpinVessel Speed 50 RPM Rotate 2 1 Pause 200 milliseconds Ramp 10 It is always a good idea to practice with water in your SpinVessel to determine safe RPM Rotation Pause and Ramp numbers before you use expensive reagents Note that if Ramp is set to 0 liquid will splash out of the Spin Vessel The V...

Page 8: ...e to comply with safety requirements and practices WARNING To avoid personnel injury caused by electrical shock do not remove the cover of the controller when the power is ON CAUTION Do not disconnect motor during operation Otherwise over current breakdown may result WARRANTY V P Scientific Inc warrants this product to be free from defects in material and workmanship when used under normal laborat...

Reviews: