background image

13

1.

  Focusing adjustment is achieved by turning the coarse/fine focus controls (see figures 19a and 20a). The large knob 

is used for coarse adjustment, the smaller knob for fine adjustment. The coaxial arrangement allows for easy, precise 

adjustment without stage drift.

2.

  Turning the coarse/fine focus control raises and lowers the stage vertically. One complete turn of the fine focusing knob 

raises or lowers the stage 0.3mm; the smallest graduation refers to 2 microns of vertical movement. One complete turn 

of the coarse focusing knob raises or lowers the stage 3.6mm. To ensure long life, turn the focusing knobs slowly and 

uniformly.

3.

  The focusing tension control knob is located just inside of the right-hand focus control knob (see figure 19b). For tighter 

tension, turn the control knob in a clockwise motion. For looser tension, turn the control knob in a counterclockwise 

motion.

4.

  Vertical Focusing: The condenser can be raised and lowered with the substage adjustment knob (see figure 20b) to 

focus the light for optimal illumination.

5.

  Aperture Adjustment: The light path can be adjusted with the iris diaphragm adjustment lever located underneath the 

condenser. Aperture adjustments are made to induce contrast into a specimen, not to adjust light intensity.

Focusing and Mechanical Stage Mechanisms

The up-stop mechanism is located just inside of the left-hand focus control knob (see figure 20c). It allows the user to set a 

maximum point to which the stage can be raised.  

Setting the Up-Stop Mechanism

1.

  To set this point, turn the up-stop mechanism in a counterclockwise motion, so that its tab is facing up.  

2.

  Raise or lower the stage, by turning the focus control knobs, to the desired height.  

Be careful not to raise the stage 

high enough to crash into the objective

3.

  Once achieved, turn the up-stop mechanism in a clockwise motion, so that its tab is facing forward.  

4.

  Once gently tightened, the up-stop mechanism will not allow the stage to be raised higher than the set point.

Figure 19

Figure 20

Using your 1400FLi Series 

Microscope

6.

  The mechanical stage X-Y controls, located underneath the right-hand side of the mechanical stage (see figure 19c), 

provide easy and accurate positioning of the sample. One complete turn of the longitudinal (Y) control (lower half of the 

stage controls) will move the specimen 34mm left or right. One complete turn of the transverse (X) control (upper half 

of the stage controls) will move the specimen 20mm front or back.

7.

  The spring-loaded slide holder can be removed for users who prefer to not use a mechanical stage. Simply loosen the 

knurled slide holder knobs which lock the slide holder on the stage, and slip out the slide holder.

[a]

[b]

[b]

[c]

[c]

All Models

All Models

Summary of Contents for 1400FLi Series

Page 1: ...Operation Manual 1400FLi Series Covering Models 1482FLi 1486FLi...

Page 2: ...adjustment and safety autostop Condenser Model 1482FLi comes with a 1 25 N A Abbe Condenser Model 1486FLi comes with a 1 25 N A Zernike condenser with phase annulus rings for 10X 20X 40X and 100X als...

Page 3: ...uction Manual 1 ea Model 1486FLi Trinocular Head 1 ea Stand 1 ea Phase Contrast Brightfield Darkfield Condenser 1 ea Phase Contrast Centering Telescope 1 ea Epi Illumination Assembly 1 ea External Pow...

Page 4: ...ra System 5 Megapixels 1400 CDPC 3 USB Color Digital Camera System 3 Megapixels Other Accessories Part Number Description Call for Part Numbers Scale Reticle Call for Part Numbers Stage Micrometer Cal...

Page 5: ...Diopter Adjustment Mechanism Interpupillary Adjustment Mechanism UV Shield Set Screw Mercury Lamp Centering Window Horizon tal Lamp Centering Knob Vertical Lamp Centering Knob Lamp Door Set Knob Epi I...

Page 6: ...rnal power supply are shown on page 16 Power Switch Variable Lighting Control Up Stop Mechanism 1400FLi Series Base Left Side View Substage Adjustment Knob Power Cord Jack 1400FLi Series Stand Rear Vi...

Page 7: ...bly on top of the epi illuminator housing assembly so that the dovetail flange slides into place Secure with the knurled head set screw upper see figure 2a NOTE Do not release the head until it is fir...

Page 8: ...holes of the nosepiece see figure 6 9 Raise the substage and stage to their maximum height Raise the stage via the coarse fine focus controls see figure 7a and the substage using the substage adjustm...

Page 9: ...bracket access door located on the right side of the housing Once loosened lift the access door up and away from the housing 14 The epi illuminator lamp bracket is shipped with a protective insulating...

Page 10: ...t of the Mercury Lamp NOTE Please see page 16 to familiarize yourself with the operation of the external power supply before turning the external power supply on 21 Plug the power supply cable see fig...

Page 11: ...kwise Closing the iris in this manner will reduce the field so that a small white hexagon is visible within a black field see figure 15A Focusing of the hexagon is performed by turning the substage ad...

Page 12: ...ce of the phase contrast centering telescope until the image is in focus Tighten the set screw 5 The image seen through the phase contrast centering telescope should resemble rings superimposed on one...

Page 13: ...intensity Focusing and Mechanical Stage Mechanisms The up stop mechanism is located just inside of the left hand focus control knob see figure 20c It allows the user to set a maximum point to which t...

Page 14: ...stment mechanism located on the right eyetube see figure 22a Once complete the microscope is corrected for your vision Interpupillary and Diopter Adjustments The 100X objective which comes with this m...

Page 15: ...Li is used mainly for adding contrast and photograph color correction during phase contrast work The blue filter is used to approximate natural light and photograph color correction Filtering is a use...

Page 16: ...replacement reset the timer by inserting a small pin paperclip etc into the reset hole located on the front of the power supply face 7 Although the lamp has a finite service life it is recommended th...

Page 17: ...ur 1400FLi Series Microscope 100W HBO Short Arc Mercury Lamps The supplied 100W HBO Short Arc Mercury Lamps have a discharge vessel made of high optical quality quartz glass containing mercury and a n...

Page 18: ...ut be careful not to scratch the circuit board 4 Replace the rear cover and the five screws Replacing the Microscope Fuse 1 Before attempting to replace or remove the lamp UNPLUG THE MICROSCOPE FROM A...

Page 19: ...0X 1 25 N A 0 25mm W D 0 18mm F O V Fluorescence Filter Cubes Blue Broadband Dichroic 500nm Excitation 430 490nm Emmision 520nm Intended Dyes FITC Acridine Orange Auramine Green Broadband Dichroic 570...

Page 20: ...ir or replacement of parts by the End User or an unauthorized third party while under warranty 4 Consumable items such as but not limited to bulbs have failed This warranty expressly excludes transpor...

Page 21: ...Printed in the USA VGMNL081319 1480FLi www veegee com 800 423 8842 sales veegee com techsupport veegee com veegee com pages technical support service...

Reviews: