Aligning the Condenser

Adjustments to the substage condensing system are crucial for proper illumination and performance. There are three basic 

adjustments which need to be made: Centering, Vertical Focusing, and Aperture Adjustment.

Figure 15A

Figure 15B

Figure 15C

Figure 15D

1.

 

Centering

: The 

condenser

 must be centered in the light path to ensure proper light control. A simple method for  

 

centering is as follows:  

 

  

  Rotate the 

nosepiece

 until the 10X 

objective

 is in the light path.

  

  Raise the substage assembly fully by turning the substage adjustment knob counter-clockwise.

 

  

  Use the focus controls to focus on a specimen located on the stage.

 

  

  Open the 

aperature iris diaphragm

 to the largest setting by using the 

aperature iris diaphragm adjustment

 lever  

 

which extends from the 

condenser

 assembly.

 
  

  While looking into the microscope eyepieces, close the 

field iris diaphragm

 to the smallest setting by turning the    

uppermost section of the 

substage illuminator

 counter-clockwise.

 

  

  Closing the iris in this manner will reduce the field so that a small white hexagon is visible within a black field (see    

figure 15A).  Focusing of the hexagon is performed by turning the 

substage adjustment knob

. This white hexagon is the    

light which is passing through the field iris and should be centered in the black field. If not, move it to the center (see  

 

figure 15B) by tightening and/or loosening the 

condenser centering knobs

.

 

  

  Fine tuning can be done by opening the 

field iris diaphragm

 until the white hexagon almost fills the entire field (see   

figure 15C), and then readjusting (see figure 15D). After centering the condenser open the field iris diaphragm slightly  

 

wider than the field of view.

If the iris isn’t centered in the field of view then use the condenser centering knobs to adjust. You may have to split the 

alignment difference between the field iris and the condenser iris to achieve a balance.

2.   Vertical Focusing

: The 

condenser

 can be raised and lowered with the 

substage adjustment

 knob to focus the light for  

 

optimal illumination.

3.   Aperture Adjustment

: The light path can be adjusted with the 

aperature iris diaphragm adjustment

 lever located just   

below the condenser. Aperture adjustments are made to induce contrast into a specimen, not to adjust light intensity.

All Models

Summary of Contents for 1400FLi Series

Page 1: ...Operation Manual 1400FLi Series Covering Models 1482FLi 1486FLi...

Page 2: ...adjustment and safety autostop Condenser Model 1482FLi comes with a 1 25 N A Abbe Condenser Model 1486FLi comes with a 1 25 N A Zernike condenser with phase annulus rings for 10X 20X 40X and 100X als...

Page 3: ...uction Manual 1 ea Model 1486FLi Trinocular Head 1 ea Stand 1 ea Phase Contrast Brightfield Darkfield Condenser 1 ea Phase Contrast Centering Telescope 1 ea Epi Illumination Assembly 1 ea External Pow...

Page 4: ...ra System 5 Megapixels 1400 CDPC 3 USB Color Digital Camera System 3 Megapixels Other Accessories Part Number Description Call for Part Numbers Scale Reticle Call for Part Numbers Stage Micrometer Cal...

Page 5: ...Diopter Adjustment Mechanism Interpupillary Adjustment Mechanism UV Shield Set Screw Mercury Lamp Centering Window Horizon tal Lamp Centering Knob Vertical Lamp Centering Knob Lamp Door Set Knob Epi I...

Page 6: ...rnal power supply are shown on page 16 Power Switch Variable Lighting Control Up Stop Mechanism 1400FLi Series Base Left Side View Substage Adjustment Knob Power Cord Jack 1400FLi Series Stand Rear Vi...

Page 7: ...bly on top of the epi illuminator housing assembly so that the dovetail flange slides into place Secure with the knurled head set screw upper see figure 2a NOTE Do not release the head until it is fir...

Page 8: ...holes of the nosepiece see figure 6 9 Raise the substage and stage to their maximum height Raise the stage via the coarse fine focus controls see figure 7a and the substage using the substage adjustm...

Page 9: ...bracket access door located on the right side of the housing Once loosened lift the access door up and away from the housing 14 The epi illuminator lamp bracket is shipped with a protective insulating...

Page 10: ...t of the Mercury Lamp NOTE Please see page 16 to familiarize yourself with the operation of the external power supply before turning the external power supply on 21 Plug the power supply cable see fig...

Page 11: ...kwise Closing the iris in this manner will reduce the field so that a small white hexagon is visible within a black field see figure 15A Focusing of the hexagon is performed by turning the substage ad...

Page 12: ...ce of the phase contrast centering telescope until the image is in focus Tighten the set screw 5 The image seen through the phase contrast centering telescope should resemble rings superimposed on one...

Page 13: ...intensity Focusing and Mechanical Stage Mechanisms The up stop mechanism is located just inside of the left hand focus control knob see figure 20c It allows the user to set a maximum point to which t...

Page 14: ...stment mechanism located on the right eyetube see figure 22a Once complete the microscope is corrected for your vision Interpupillary and Diopter Adjustments The 100X objective which comes with this m...

Page 15: ...Li is used mainly for adding contrast and photograph color correction during phase contrast work The blue filter is used to approximate natural light and photograph color correction Filtering is a use...

Page 16: ...replacement reset the timer by inserting a small pin paperclip etc into the reset hole located on the front of the power supply face 7 Although the lamp has a finite service life it is recommended th...

Page 17: ...ur 1400FLi Series Microscope 100W HBO Short Arc Mercury Lamps The supplied 100W HBO Short Arc Mercury Lamps have a discharge vessel made of high optical quality quartz glass containing mercury and a n...

Page 18: ...ut be careful not to scratch the circuit board 4 Replace the rear cover and the five screws Replacing the Microscope Fuse 1 Before attempting to replace or remove the lamp UNPLUG THE MICROSCOPE FROM A...

Page 19: ...0X 1 25 N A 0 25mm W D 0 18mm F O V Fluorescence Filter Cubes Blue Broadband Dichroic 500nm Excitation 430 490nm Emmision 520nm Intended Dyes FITC Acridine Orange Auramine Green Broadband Dichroic 570...

Page 20: ...ir or replacement of parts by the End User or an unauthorized third party while under warranty 4 Consumable items such as but not limited to bulbs have failed This warranty expressly excludes transpor...

Page 21: ...Printed in the USA VGMNL081319 1480FLi www veegee com 800 423 8842 sales veegee com techsupport veegee com veegee com pages technical support service...

Reviews: