2.3 Probe parts
shows the MGP260 main components with the connection box of the
probe closed and opened.
1
2
3
4
5
6
8
9
10
11
7
Figure 2 MGP260 probe parts (closed and opened view)
1
Grounding terminal: use a
≥
4 mm
2
wire to connect to the grounding rail
2
Breather plug and lead-throughs for wiring: install cable glands as required and seal
unused lead-throughs (see
Lead-through options and breather plug (page 13)
)
3
Tightening flange of the connection box with captive Allen screws. Loosen the captive
screws with a 5 mm (3/16 in) Allen key and open the connection box to access wiring
terminals.
4
Tightening nut: only tighten from the tightening nut when installing (wrench size: 50 mm
(1.97 in))
5
1.5" male NPT thread:
never install to any other thread type than 1.5" female NPT thread
6
Probe filter (user-replaceable)
7
1.5" NPT thread test plug
8
Wiring terminals for optional 4 … 20 mA input from external pressure or temperature
sensor (Ex ia)
9
Barrier separating the intrinsically safe (Ex ia) optional external sensor input terminals
from the analog output, power supply input, and RS-485 terminals
10
Wiring terminals for 4 … 20 mA analog outputs, 18 … 30 VDC power supply input, and
RS-485 communication
11
Measurement cuvette with optics and CARBOCAP
â
sensor inside the probe filter
MGP260 User Guide
M212504EN-A
12
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备 气象科学仪器设备
Tel:029-88186182 Web:www.xavery.cn Email:[email protected]
Summary of Contents for MGP260 Series
Page 8: ...MGP260 User Guide M212504EN A 6 Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Page 88: ...Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Page 90: ...Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Page 91: ...Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Page 92: ...www vaisala com Tel 029 88186182 Web www xavery cn Email sxvery 163 com...