Analog output 3
2309
0904
hex
Scale high end for analog
output 3 measurement
parameter.
Minimum and maximum
values vary for different
parameters.
Float
Output
parameters
0000
hex
(CH
4
wet basis)
0002
hex
(CH
4
dry basis)
0004
hex
(CO
2
wet basis)
0006
hex
(CO
2
dry basis):
Minimum: 0
Maximum:
1000000
Output
parameters
0008
hex
(H
2
O
wet basis)
000A
hex
(H
2
O
dry basis):
Minimum: 0
Maximum:
250000
Output
parameters
000C
hex
(T
d
)
000E
hex
(T
df
):
Minimum: -10
Maximum: 60
2311
0906
hex
Error output level (mA)
Float
Min. 0.5 mA
Max. 24 mA
Analog input 1
2562
0A01
hex
Analog input 1
compensation parameter.
Read-only: controlled by
the pressure and
temperature
compensation mode
registers.
Reg
0204
hex
(pressure)
0206
hex
(temperature)
Appendix A – Modbus reference