6. Attach the mounting plate to the mounting surface with screws or ties. For mounting
plate dimensions, see
7. Check that the probe sits closely in the clamps after mounting. The mounting kit and the
flow-through adapter have been aligned so that the probe sits in the adapter in a straight
angle. A gap between the probe body and clamp indicates that the mounting alignment is
incorrect and must be adjusted.
90°
Ensure the adapter mounting does not pull the probe out of a
straight angle. If the mounting pulls the probe to the side, the leak
tightness of the gland O-ring can be affected.
CAUTION!
8. Connect the gas inlet tube to the upper port of the adapter and the gas outlet tube to the
lower port. Adapter port size: G 1/8.
More information
‣
Installation option: flow-through adapter installation (page 22)
‣
Flow-through adapter dimensions (page 66)
Chapter 3 – Installation
33