1. Loosen the 2 screws on the clamp.
2. Loosen the probe gland.
Only rotate the tightening nut of the probe gland. Never turn
from the locknut of the probe gland (below the tightening nut). The
threads of the locknut are glued onto the adapter with a sealant, and
rotating the locknut will compromise the leak tightness of the adapter.
CAUTION!
3. Insert the probe head into the adapter through the clamp and probe gland. Push the
probe into the adapter until the shoulder of the probe head meets the slot inside the
adapter and stops. The probe filter should cover both the inlet and outlet ports from the
inside. Check the installation depth measurement as shown in
4. Tighten the probe gland.
5. After tightening the probe gland, tighten the 2 screws on the clamp.
A high process pressure can cause the probe to detach from
the adapter if the clamp and gland have been left loose. Always fully
tighten the clamp and probe gland.
WARNING!
MGP260 User Guide
M212504EN-A