2.3.1 Lead-through options and breather plug
shows the MGP260 lead-through options and breather plug location.
Cable glands are not provided by Vaisala. When selecting cable glands for your
application, note the requirements in
Guidelines for safe use in hazardous
1
3
2
4
Figure 3 Lead-through options and breather plug
1
Breather plug: do not adjust.
2
M16x1.5 wiring lead-through: used when wiring the optional Ex ia external pressure or
temperature sensor input terminal.
3
M20x1.5 wiring lead-through with removable plastic transport cover (fold sides of
transport cover together and pull out). Wiring option for analog output, power supply
input, and RS-485 terminals, or customer-specific additional wiring.
4
M20x1.5 wiring lead-through for analog output, power supply input, and RS-485 terminals.
Seal unused lead-throughs. Use only Ex compliant seals on the
MGP260 lead-throughs.
CAUTION!
More information
‣
Preparing probe for installation (page 26)
‣
Chapter 2 – Product overview
13