PR-23 instruction manual
8.1.7 Message
HIGH SENSOR HUMIDITY
Tells that humidity measured at the Sensor processor card exceeds 60 % relative hu-
midity. The reason may be moisture leaking in through prism seal or the cover being
open. Also, check and, if necessary, replace prism seal.
8.1.8 Message
HIGH SENSOR TEMP
The temperature on the Sensor processor card exceeds 65 °C (150 °F). To read this
temperature, select
3 SENSOR STATUS
from the Main menu. For action, see Section 2.2.1,
“Choosing sensor mounting location”.
8.1.9 Message
HIGH TRANSMITTER TEMP
The temperature of the Motherboard of the Indicating transmitter exceeds 60 °C
(140 °F). To read this temperature, select
3 SENSOR STATUS
from the Main menu and
check
DTR TMP
. If the warning persists, the transmitter should be moved to a cooler
place (for example out of sun).
8.1.10 Message
LOW TRANSMITTER VOLT
The internal DC voltages of the transmitter are below specifications. Check the power
supply input voltage. If the supply voltage is within specifications, replace power sup-
ply module, Figure 8.1.
8.1.11 Relays and switches not working
Check configuration, Section 5.3, “Viewing system information”, and for possible cor-
rection Section 6.3.2, “Configuring input switches”, Section 6.3.1 “Configuring relays”,
and Section 6.5 “Configuring prism wash”.
Relay status is indicated by LEDs D23, D25 on the Motherboard, see Section 8.1.2 “Di-
agnostic LEDs”. For switches, check also LED D26 on the Motherboard indicating that
the 3 V DC supply is correct, see Section 8.1.2 “Diagnostic LEDs”.
The wash function can be tested according to Section 5.1.3 “Testing prism wash”.
8.1.12 Output signal error during
NORMAL OPERATION
If there is no output signal, check wiring (Section 3.3, “Electrical connections”) and
the Diagnostic LED D26 (Section 8.1.2, “Diagnostic LEDs”).
If the mA signal does not correspond to the concentration display, check output signal
configuration, Section 5.3, “Viewing system information”, and for possible correction
Section 6.3.3, “Configuring mA outputs”. A low mA signal can also be caused by high
resistance in the external current loop, see Section 3.3, “Electrical connections”.
A noisy signal can be damped, Section 6.1, “Configuring output signal damping”.
Summary of Contents for K-PATENTS PR-23 AX Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...