8 Troubleshooting
69
8.1.4 Message
NO SENSOR
Cause
: The current in the cable to this sensor is below 20 mA. Normally this means
that there’s no sensor connected to the cable or that there’s no cable to the DTR. If
this message comes up while a sensor properly is connected, the most likely cause of
this message is a fault in the sensor. It is also possible that the cable is totally dead for
example if it is accidentally cut through.
See also Diagnostic LED LD1/LD3, Section 8.1.2 “Diagnostic LEDs”. The concentration
display will be a dashed line.
8.1.5 Message
NO SIGNAL
Besides the message, the concentration display will be a dashed line although a sensor
is connected.
Cause
: The current in the cable to this sensor is in the correct range 20–60 mA, but
no data is coming in from the sensor. This indicates that the Sensor processor card is
faulty.
See also Diagnostic LED LD1/LD3, Section 8.1.2 “Diagnostic LEDs”.
Action
: Replace the Sensor processor card.
8.1.6 Message
SHORT-CIRCUIT
The current in the cable to the sensor A/B exceeds 60 mA. First, the DTR attempts
for a short time to reconnect with the sensor in question. If the short-circuit persists,
the sensor in question is switched off completely to protect the Motherboard from
overheating.
See also Diagnostic LED LD2/LD4, Section 8.1.2, “Diagnostic LEDs”.
Note:
If two sensors are connected to the DTR, a short-circuit in one of the cables may
disturb the measurement of both sensors as DTR attempts to reconnect. The mea-
surement of the non-affected sensor returns to normal as soon as the short-circuited
sensor is switched off.
If the DTR detects a short-circuit that persists, the affected sensor is switched off to
prevent further damage. The message
SHORT-CIRCUIT
will stay on the screen until the
DTR is powered off and on.
See also Diagnostic LED LD2/LD4, Section 8.1.2, “Diagnostic LEDs”.
Cause and action
: The most likely cause of these messages is a problem in the cable
connecting the sensor in question to the DTR. Check that the cable is undamaged and
replace it if necessary, then turn the DTR off and back on.
Summary of Contents for K-PATENTS PR-23 AX Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...