PR-23 instruction manual
6.2.3 Tolerance time
The tolerance time setting can be used in processes where there are intermittent
breaks in the measurement due to non-representative sample on the prism. This typ-
ically occurs when there are large voids in the process liquid.
If the optical image can be interpreted, the tolerance time setting does not have any
effect. When the optical image can no longer be interpreted (status messages, e.g.,
NO SAMPLE
,
NO OPTICAL IMAGE
,
PRISM COATED
), the measurement is held for the given
number of seconds.
For example, a setting of 10 seconds ensures that any
NO SAMPLE
state which is shorter
than 10 seconds will not make a dip into the output signal. The factory setting is 5
seconds, access the
6 TOLERANCE TIME
menu item to set the tolerance time.
The tolerance time counter is reset always when there is a representative sample on
the prism (e.g., the nD can be determined). Figure 6.4 illustrates this behaviour with
an intermittent measurement signal. When the signal drop is shorter than the toler-
ance time (e.g. at t = 10 s or t = 35 s in the figure), the output signal does not drop. If
the signal drop is so long that the tolerance time counter reaches zero, there will be a
drop in the output signal (at t = 80 s in the figure).
CONC [%]
Signal with tolerance
Original signal
Filtered (damped) output
0
10
20
30
40
50
60
70
80
90
100
Time [s]
Tol. time counter
Figure 6.4
Effect of tolerance time on output
6.2.4 QF threshold
The QF threshold setting can be used to prevent the instrument from measuring when
the image quality is below a certain limiting value. When QF value becomes less than
the user defined value the image status changes to
NO OPTICAL IMAGE
after the user
Summary of Contents for K-PATENTS PR-23 AX Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...