6 Configuration and calibration
55
6.5.1 Wash cycle
The wash logic is shown in Figure 6.15 as a flow diagram. The automatic prism wash
cycle (Figure 6.14) consists of three phases:
precondition
,
wash
and
recovery
. The op-
tional preconditioning function is used to e.g. blow out the condensate before wash-
ing. After the preconditioning there is a one-second pause to avoid having both pre-
condition and wash relays active at the same time.
The wash cycle is initiated when the wash interval has elapsed. The wash can also be
started by closing an external switch (Remote wash, see Section 6.3.2) or manually
from the user interface at the Sensor status display (see Section 5.1.3). The order of
priority for these wash triggers is:
1. manual wash
2. remote wash request
3. wash interval timer
WASH
RECOVERY
INTERVAL
PRECOND.
Figure 6.14
The automatic prism wash cycle
Note:
For safety reasons two sensors never wash simultaneously. If the manual wash
button for sensor A is pressed while sensor B is washing, the wash cycle for sensor A
is started after B has finished. Similarly, if the interval time for sensor B elapses when
A is washing, the wash for sensor B is delayed until A has finished.
In case of remote wash request the request is discarded if it arrives when the other
sensor is washing. The request is honored only if the contacts are held closed until
the other sensor has finished.
The wash relay is closed for the wash time specified in the wash settings. If the wash
auto-cut functionality is active, the wash may be ended earlier (see Figure 6.16). The
specified wash time is never exceeded.
After the wash phase is completed, a recovery time is spent. During the wash cycle
(precondition, wash, recovery) the measurement result is in hold unless otherwise
specified.
Summary of Contents for K-PATENTS PR-23 AX Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...