10
PR-23 instruction manual
3.2 Mounting Indicating transmitter
The Indicating transmitter is mounted indoors, it should preferably be located in an
easily accessible, well lit and dry area. Avoid vibration. Take interconnecting cable
length into consideration when choosing the mounting location.
The enclosure is mounted vertically on an upright surface (wall) using four mounting
feet, see Figure 3.3. The LCD is best viewed when approximately on the eye level of
the user.
In sanitary installations, the recommendation is to use a DTR with stainless steel en-
closure. If standard polycarbonate enclosure is used, it should be installed as remotely
as practical from the product areas or connections.
Important:
Do not drill mounting holes in the enclosure as that will affect the pro-
tection class of the enclosure and damage the electronics.
Figure 3.3
Indicating transmitter: dimensions (mm/in)
and mounting feet measures
Note:
The LCD display has an operating temperature range of 0–50 °C and a storage
temperature range of -20–60 °C.
Important:
The DTR does not have a built-in power switch. The system is always
powered on when connected to a power source. Mounting an external power switch
to control the DTR’s power supply is recommended, Figure 3.6.
3.3 Electrical connections
3.3.1 Interconnecting cable
The cable contains a pair of twisted signal wires (1, 2) and a cable shield (3) (see
Section 3.3.2 and Figure 3.7). Standard delivery is 10 meters (33 feet) of cable. The
maximum length of an interconnecting cable is 200 m (660 ft). The signal wires (1, 2)
are interchangeable (non-polarized). The cable shield is connected to the protective
earth at the Indicating transmitter.
Summary of Contents for K-PATENTS PR-23 AX Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...