1.35
%
| Carbon dioxide
WLAN is activated
Indigo 201
3
2
1
Figure 9 Attaching probes and cables to Indigo200
1
Probes are locked in place with the locking wheel.
Never turn from the probe body.
2
Connect probe cables in the same way as probes: insert the cable in the connector and
hold in place while turning the locking wheel.
3
Insert probes into the probe connector with the orientation mark facing out.
1. Insert the probe into the probe connector with the orientation mark on the probe body
facing out.
2. Hold the probe in the probe connector and lock it in place by turning the locking wheel
counterclockwise. Do not turn the probe body when attaching, only the locking wheel on
the transmitter.
3. If no previous configuration exists (that is, this is the first time a probe is connected to
Indigo, and no configuration has been set manually), Indigo adapts the analog output
settings of the connected Vaisala Indigo-compatible probe automatically.
If a configuration that differs from the attached probe's configuration exists
in the transmitter, you must configure the new probe manually to enable
analog output. See
Analog output configuration overview (page 28)
Indigo201 User Guide
M212842EN-B
20