Figure 26 Diagnostics Tab, CO
2
Probe Example
7.5 Configuration Tab
The
Configuration
tab of the
Calibration
menu contains probe-specific configuration
options and the environmental compensation options.
Note that the probe configuration options available in this view are not always necessary for
use with Indigo. For example, the
Communication
options shown in the following CO
2
probe
example apply only to the probe's own communication settings. The communication
settings of the Indigo transmitter are configured in a separate menu.
Similarly, the analog output settings shown here apply only for the probe's analog output
channels, which are not used when the probe is connected to an Indigo transmitter that has
its own output channels and settings.
Chapter 7 – Calibration and Adjustment
51
Summary of Contents for Indigo 202
Page 1: ...M211966EN A User Guide Indigo 202 Digital Transmitter Indigo 202...
Page 22: ...Indigo 202 User Guide M211966EN A 20...
Page 62: ...Indigo 202 User Guide M211966EN A 60...
Page 70: ...Indigo 202 User Guide M211966EN A 68...
Page 74: ...Indigo 202 User Guide M211966EN A 72...
Page 75: ......
Page 76: ...www vaisala com...