Table 5 Indigo Transmitter Messages Shown on Display
Error Message
Description
Recommended Action
Errors
Low supply voltage
The supply voltage is below the
minimum (range: 15 ... 30 VDC or
20 ... 22 VAC).
The error clears when the supply
voltage returns to the specified
range. Check your power supply
and restart Indigo if needed.
No legal measurement probe
The connected probe is not
compatible with Indigo or a
restriction (probe name or serial
number) on allowed probes is in
place.
Check your probe's Indigo
compatibility and possible
restrictions on allowed probes.
WLAN error
There is a problem with the
wireless connection.
Go through the troubleshooting
instructions in
Warnings
No measurement probe
The measurement probe is not
connected, or the probe is
connected poorly and cannot be
recognized by Indigo.
Check that you have fastened the
probe correctly with the locking
wheel.
Notifications
Probe connected:
Indigo displays the name of the
connected probe.
WLAN on
The WLAN connection is active
and you can connect to the
wireless configuration interface.