HMI41 AND PROBES
M210316EN-A
Operating Manual
7
4.3
HOLD-mode
In any of the previously mentioned measurement reading displays, the button
HOLD freezes the display to show the current readings, e.g. RH & T readings:
RH
hold
T
°C
%
With MODE or ENTER, the display returns to the normal display mode.
If the indicator turns itself off with the automatic power-off function during
the HOLD-mode, it wakes up in the same mode when turned on again. The
'hold'-text is blinking and the indicator can be returned to the normal display
mode with any button except ON/OFF.
4.3.1
MIN-mode
If you wish to know the minimum readings that have been measured after
power up, press HOLD when the indicator is in the HOLD-mode. The
minimum readings of the currently displayed quantities are shown (if the
‘hold’-text is blinking, the HMI41 must first be returned to the normal display
mode in order to activate the MIN-mode; see Chapter 4.3):
RH
min
T
°C
%
With MODE or ENTER, the indicator returns to the normal display mode.
4.3.2
MAX-mode
If you wish to know the maximum readings that have been measured after
power up, press HOLD when the indicator is in the MIN-mode. The maximum
readings of the currently displayed quantities appear:
RH
max
T
°C
%
The indicator returns to the normal display mode with any button except
ON/OFF.
Summary of Contents for HUMICAP HMP41
Page 1: ...USER S GUIDE Vaisala HUMICAP Indicator HMI41 and Probes HMP41 45 46 M210316EN A ...
Page 70: ......
Page 71: ......
Page 72: ...www vaisala com M210316EN ...