HMI41 AND PROBES
M210316EN-A
Operating Manual
45
Accuracy of mixing ratio (g/kg)
with
ambient pressure of 1013.25 mbar
RH/%
10
20
30
40
50
60
70
80
90
100
T/
°
C
-40
0.002 0.002 0.002 0.002 0.003 0.003 0.003 0.003
-
-
-20
0.014 0.015 0.017 0.018 0.019 0.020 0.022 0.023
-
-
0
0.08
0.09
0.09
0.10
0.10
0.11
0.12
0.12
0.13
0.13
20
0.31
0.33
0.35
0.37
0.39
0.41
0.43
0.45
0.47
0.49
40
0.97
1.03
1.10
1.17
1.24
1.31
1.38
1.46
1.54
1.62
60
2.70
2.94
3.46
3.76
3.72
4.08
4.42
4.79
5.19
5.63
80
6.78
7.80
9.00
10. 4
12.2
14.3
16.9
20.2
24.4
29.7
100
16.4
21.6
29.2
41.3
62.0
101
190
462
-
-
120
41.2
75.7
176
-
-
-
-
-
-
-
140
-
-
-
-
-
-
-
-
-
-
160
-
-
-
-
-
-
-
-
-
-
Accuracy of absolute humidity
(g/m
3
)
RH/%
10
20
30
40
50
60
70
80
90
100
T/
°
C
-40
0.003 0.003 0.003 0.003 0.004 0.004 0.004 0.005
-
-
-20
0.020 0.021 0.023 0.025 0.026 0.028 0.029 0.031
-
-
0
0.10
0.11
0.12
0.13
0.13
0.14
0.15
0.15
0.16
0.17
20
0.37
0.39
0.41
0.43
0.45
0.47
0.49
0.51
0.53
0.55
40
1.08
1.13
1.18
1.24
1.29
1.34
1.39
1.44
1.49
1.54
60
2.73
2.84
2.95
3.07
3.18
3.29
3.40
3.52
3.63
3.74
80
6.08
6.30
6.51
6.73
6.95
7.17
7.39
7.61
7.83
8.05
100
12.2
12.6
13.0
13.4
13.8
14.2
14.6
15.0
15.3
15.7
120
22.6
23.3
23.9
24.6
25.2
25.8
26.5
27.1
27.8
28.4
140
39.1
40.0
41.0
42.0
43.0
44.0
45.0
45.9
46.9
47.9
160
63.5
64.9
66.4
67.8
69.2
70.7
72.1
73.5
75.0
76.4
Accuracy of wet bulb temperature (°C)
RH/%
10
20
30
40
50
60
70
80
90
100
T/
°
C
-40
0.20
0.20
0.20
0.20
0.20
0.20
0.20
0.20
-
-
-20
0.21
0.21
0.21
0.21
0.22
0.22
0.22
0.22
-
-
0
0.27
0.28
0.28
0.29
0.29
0.29
0.30
0.30
0.31
0.31
20
0.45
0.45
0.45
0.44
0.44
0.44
0.43
0.43
0.42
0.42
40
0.84
0.77
0.72
0.67
0.64
0.61
0.58
0.56
0.54
0.52
60
1.45
1.20
1.03
0.91
0.83
0.76
0.71
0.67
0.63
0.61
80
2.24
1.64
1.32
1.13
0.99
0.90
0.82
0.76
0.72
0.68
100
3.06
2.04
1.58
1.31
1.14
1.01
0.92
0.85
0.80
0.75
120
3.86
2.41
1.81
1.48
1.28
1.13
1.03
0.95
0.88
0.83
140
4.57
2.73
2.03
1.65
1.41
1.25
1.13
1.04
0.97
0.91
160
5.23
3.04
2.24
1.81
1.55
1.36
1.23
1.13
1.05
0.98
Summary of Contents for HUMICAP HMP41
Page 1: ...USER S GUIDE Vaisala HUMICAP Indicator HMI41 and Probes HMP41 45 46 M210316EN A ...
Page 70: ......
Page 71: ......
Page 72: ...www vaisala com M210316EN ...