HMI41 AND PROBES
Operating Manual
M210316EN-A
42
11.2
HMP41and HMP45 probes
11.2.1
Relative humidity
Measurement range
0...100 %RH non-condensing
Accuracy (at +20 °C)
when calibrated against
salt solutions (ASTM E104-85):
±
2 %RH (0...90 %RH)
±
3 %RH (90...100 %RH)
Temperature dependence
of electronics
±0.05 %RH/
°
C
Typical long-term stability
better than 1 %RH per year
Response time (90%) at 20
°
C
in still air with sintered filter
15 s
Humidity sensor
HUMICAP
®
180
11.2.2
Temperature
Measurement range
(for which accuracy is specified):
-20...+60
°
C
Temperature sensor
Pt 1000 (IEC 751 1/3 Class B)
Accuracy:
°C
0.4
0.3
0.2
0.1
0
-0.1
-0.2
-0.3
-0.4
-20
0
20
40
60
°C
-10
10
30
50
11.2.3
General
Typical ranges of calculated variables
dewpoint temperature
-40...+60
°
C
absolute humidity
0...160 g/m
3
wet bulb temperature
0...+60
°
C
mixing ratio
0...160 g/kg d.a.
Cable length (HMP45)
1500 mm; extended spiral cable
Connector type (HMP45)
modular connector
Operating temperature range
-40...+60
°
C
Storage temperature range
-40...+70
°
C
Housing material
ABS plastic
Housing classification (electronics)
IP65 (NEMA 4)
Sensor protection
plastic grid, part no. HM46717
Weight:
HMP41
30 g
HMP45
160 g
Summary of Contents for HUMICAP HMP41
Page 1: ...USER S GUIDE Vaisala HUMICAP Indicator HMI41 and Probes HMP41 45 46 M210316EN A ...
Page 70: ......
Page 71: ......
Page 72: ...www vaisala com M210316EN ...