HMI41 AND PROBES
M210316EN-A
Operating Manual
31
>
To change the output, type
CPLAY
, then the character you wish to appear
between decimals, then the character you wish to use between fields and then
<cr>. An example:
>cplay , <cr>
Desimal separator : ,
Field separator : TAB
example:
1 01:00:00 38,72 21,61 7,01
>
7.2.3
HELP
Outputting available commands and their
contents
If you wish to see which commands are available, type HELP and press
ENTER. The following list appears:
>help
Available commands :
HELP ? PLAY CPLAY
Type HELP <command_name> for more help
>
To see the contents of each command, type
HELP
, command name (e.g.
PLAY
)
and press ENTER; this brings an explanation of the command and its usage on
the display:
>help play
Command : PLAY
Purpose : Send recordings from memory to serial port
Usage : PLAY hh:mm <cr>, hh:mm = rec starting time (optional)
if command is used without parameters it uses default setting
>
7.3.4
?
Outputting the HMI41 settings
If you need to know which parameters and settings are currently stored in your
HMI41 indicator, type ? and press ENTER:
>?
HMI41 / 2.01
Serial number : A0000000
Output units : metric
Baud P D S : 4800 E 7 1 FDX
Pressure : 1013.25
Auto Off : 5
Probe : 2
Start-up mode : 1
4.th variable : none
>