6. Connect the wires of the connection cable. When using cables provided with DMT132,
refer to
Wiring (page 19)
. See also
Power Supply Requirements (page 20)
.
Figure 13 Connecting the Cable and LED Indicator Plug
7. Plug in the cable to connector I of the transmitter and tighten it. For connector pinout,
see
Wiring (page 19)
.
8. If you have the LED indicator plug option, plug it in to connector II of the transmitter
and tighten it.
9. Turn on the power supply to start up the DMT132. The analog output will stabilize in five
seconds after start-up.
3.3 Wiring
The power supply lines on connector II are intended for use by the USB
service cable and the MI70 hand-held indicator. Do not connect other power supplies to
connector II.
CAUTION!
Table 4 Connector Pinouts
Pin
Connector I
Connector II
Wire color
1
VDC
VDC
Brown
2
Not connected
RS-485 D0-
White
3
GND
GND
Blue
4
Not connected
RS-485 D1+
Black