Command
Description
FRESTORE
Restore factory settings
HELP
Display the command list
L
View user adjustment parameters
LI
Set user adjustment parameters
PASS
Access advanced service commands. Requires passcode. Use when instructed by
Vaisala personnel.
PRES
Set value for pressure compensation
RHCAL
Calibrate humidity measurement
RESET
Reset the transmitter
SEND
Output the reading once
SYSTEM
Show transmitter software information
VERS
Show transmitter firmware version
4.4 Device Information and Status
4.4.1 Show Device Information
The
?
command outputs a listing of device information.
?
<cr>
Example:
>
?
>Device Name : DMT132 Dewpoint transmitter
Calibrated : "20110223"
Calibration location: "VAISALA/HEL"
Serial number : "G0831004"
Batch number : "G0000007"
Sensor model : "HUMICAP180R"
Sensor serial number: "F312"
Operating mode : 2 Wire
Transmitter status : NORMAL