Allow sufficient cooling of the pump before
starting maintenance.
+
Vent the pump and isolate it from the
vacuum system before you start mainte-
nance.
Ensure that maintenance is done only by suitably
trained and supervised technicians.
The valves and diaphragms as well as the motor capac-
itors are wear parts. If the rated ultimate vacuum is no
longer achieved or in case of increased noise level, the
pump interior, the diaphragms and the valves must be
cleaned and the diaphragms and valves must be checked
for cracks or other damage.
All bearings are encapsulated and are filled with long-life
lubricant. Under normal operating conditions, the drive
system is maintenance free.
In demanding circumstances, it may be efficient to check
and clean the pump heads on a regular basis. In normal
use, the lifetime of the diaphragms and valves is typically
15,000 operating hours.
- Prevent internal condensation, transfer of liquids or
dust. The diaphragms and valves will be damaged if
liquid is pumped in significant amount.
- Carry out maintenance frequently if the pump is ex-
posed to corrosive media or in case of deposits.
- Regular maintenance will improve the lifetime of the
pump and also protect both users and the environment.
NOTICE
atm
°C / °F
+
Service only one side of the pump at a time to avoid the mixing of parts.
+
In case, support the pump appropriately.
Tools required (metric):
- Torx driver T10 / T20
- 4 / 5 mm wide Allen key
- 2.5 mm wide slotted screwdriver
- Flat pliers
- Phillips screwdriver size 2
- Diaphragm key width 66 mm
Summary of Contents for ME 16C NT-ME 16C NT+EK
Page 29: ...page 29 of 88 ME 16C NT MD 12C NT MV 10C NT 1 5 3 2 6 4 8 7 9...
Page 49: ...page 49 of 88 Disassembling the pump housing 1 3 2...
Page 50: ...page 50 of 88 4 6 5 1 2 4 4x 2 5 mm 2 5 mm...
Page 51: ...page 51 of 88 7 9 8 4 4x without EK...
Page 52: ...page 52 of 88 10 12 11 4 2x 4 4x Don t loosen this screw...
Page 53: ...page 53 of 88 13 15 14 5 2x 5 2x...
Page 54: ...page 54 of 88 16 18 17 2 5 mm...
Page 56: ...page 56 of 88 Replacing the diaphragms 19 21 20 Clean 5 8x 20696867 20696839...
Page 57: ...page 57 of 88 22 24 23 1 2 3 Pay attention to number of washers...
Page 58: ...page 58 of 88 25 27 26...
Page 61: ...page 61 of 88 34 36 35 Clean 20696867 20696839...
Page 62: ...page 62 of 88 37 39 38 IN IN EX EX 4x...
Page 63: ...page 63 of 88 40 42 41 IN EX T20 4x 2 2 ft lbf 3 Nm...
Page 64: ...page 64 of 88 43 45A 44 Maintain all pump heads in the same way MD 12C NT...
Page 65: ...page 65 of 88 45B 45C 45D ME 16C NT MV 10C NT ME 16C NT MD 12C NT MV 10C NT...
Page 67: ...page 67 of 88 4A 3 2 size 2 2x Version A valve 24 4 mm...
Page 68: ...page 68 of 88 5A 20696839 24 4 mm 26 mm 4B 5B 20696839 37 mm 40 mm Version B valve 37 mm...
Page 69: ...page 69 of 88 6 7 size 2 2x 7A without EK size 2 2x...
Page 71: ...page 71 of 88 Assembling the pump housing 46 47 48 5 2x 8 9 ft lbf 12 Nm...
Page 72: ...page 72 of 88 50 49 5 2x 8 9 ft lbf 12 Nm 51...
Page 73: ...page 73 of 88 53 52 4 4x 3 7 ft lbf 5 Nm 54 4 2x 3 7 ft lbf 5 Nm...
Page 74: ...page 74 of 88 56 55 4 4x 3 7 ft lbf 5 Nm 57...
Page 75: ...page 75 of 88 59 58 60 4 4x 3 7 ft lbf 5 Nm...
Page 76: ...page 76 of 88 62 61 63...
Page 79: ...page 79 of 88 4 20638821 20638821 5 6...
Page 80: ...page 80 of 88 7 T20 4x size 1 4x 8...
Page 83: ...page 83 of 88 Declarations and certificates...
Page 84: ...page 84 of 88...
Page 87: ...page 87 of 88...